Abstract

The capability of error detection of patient-specific QA tools plays an important role in verifying MLC motion accuracy. The goal of this study was to investigate the capability in error detection of portal dosimetry, MapCHECK2 and MatriXX QA tools in IMRT plans. The 9 fields IMRT for 4 head and neck plans and 7 fields IMRT for 4 prostate plans were selected for the error detection of QA devices. The measurements were undertaken for the original plan and the modified plans, where the known errors were introduced for increasing and decreasing of prescribed dose (±2%, ±4% and ±6%) and position shifted in X-axis and Y-axis (±1, ±2, ±3 and ±5 mm). After measurement, the results were compared between calculated and measured values using gamma analysis at 3%/3 mm criteria. The average gamma pass for no errors introduced in head and neck plans was 96.9%, 98.6%, and 98.8%, while prostate plans presented 99.4%, 99.0%, and 99.7%, for portal dosimetry, MapCHECK2 and MatriXX system, respectively. In head and neck plan, the shifted error detections were 1 mm for portal dosimetry, 2 mm for MapCHECK2, and 3 mm for MatriXX system. In prostate plan, the shifted error detections were 2 mm for portal dosimetry, 3 mm for MapCHECK2, and 5 mm for MatriXX system. For the dose error detection, the portal dosimetry system could detect at 2% dose deviation in head and neck and 4% in prostate plans, while other two devices could detect at 4% dose deviation in both head and neck and prostate plans. Portal dosimetry shows slightly more capability to detect the error compared with MapCHECK2 and MatriXX system, especially in the complicated plan. It may be due to higher resolution of the detector; however, all three-detector types can detect various errors and can be used for patient-specific IMRT QA.

Highlights

  • Intensity modulated radiation therapy (IMRT) technique is one of the advanced radiation treatment techniques, which allows conformal shaping of dose distribution to tumors and significantly gains in the sparing of normal surrounding tissues

  • electronic portal imaging devices (EPID) is originally designed for treatment field verification before the treatment; it has been applied for dosimetry device as a patient-specific IMRT quality assurance (QA) due to the high spatial resolution of amorphous silicon diode array

  • The measurements were undertaken by measuring the original plan as the first step using Portal dosimetry, MapCHECK2 and MatriXX systems

Read more

Summary

Introduction

Intensity modulated radiation therapy (IMRT) technique is one of the advanced radiation treatment techniques, which allows conformal shaping of dose distribution to tumors and significantly gains in the sparing of normal surrounding tissues. IMRT achieves desired dose distribution in a complex shaped volume by modulating the intensity map of each treatment field using the moving of multileaf collimator (MLC). The position of MLC pattern needs to be verified as the patient-specific quality assurance (QA) for all IMRT plans before deliver dose to patients [1]. There are many methods to verify the patient-specific IMRT plan, such as point dose, planar dose and volume dose verification types. The example of 2D ionization array is MatriXX that consists of many vented ionization chambers. EPID is originally designed for treatment field verification before the treatment; it has been applied for dosimetry device as a patient-specific IMRT QA due to the high spatial resolution of amorphous silicon (aSi) diode array. The sensitivity of each QA tool depends on the type and number of detectors, arrangement, and spacing among them

Objectives
Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.