Abstract

An energy-dispersive (ED) X-ray computed tomography (CT) system is useful for carrying out monochromatic imaging by selecting optimal energy photons. CT is performed by repeated linear scans and rotations of an object. X-ray photons from the object are detected by the cadmium telluride (CdTe) detector, and event pulses of X-ray photons are produced using charge-sensitive and shaping amplifiers. The lower photon energy is determined by a comparator, and the maximum photon energy of 70 keV corresponds to the tube voltage. Logical pulses from the comparator are counted by a counter card through a differentiator to reduce pulse width and rise time. In the ED-CT system, tube voltage and current were 70 kV and 0.30 mA, respectively, and X-ray intensity was 18.2 µGy/s at 1.0 m from the source at a tube voltage of 70 kV. Demonstration of gadolinium K-edge CT for cancer diagnosis was carried out by selecting photons with energies ranging from 50.4 to 70 keV, and photon-count energy subtraction imaging from 30 to 50.3 keV was also performed.