Abstract

Many researches have been performed about Lab-On-Chip microfluidic systems experimentally and numerically. At the best of our knowledge, there are so few numerical assessments about Lab-On-Disk. Hence, aim of the present paper is numerical investigation of Lab-On-Disk (Laboratory-On-Disk or LOD) platforms because of its broadened usage in variety of assays in different fields of science particularly in biomedicine, drug delivery, biomedical and diagnostic assays. To fulfill this purpose, Lattice Boltzmann Method (LBM) has been employed to assess droplet generation in a T-junction through investigation of influential parameters such as Bond number (Bo), density ratio, viscosity ratio and some geometrical parameters such as width ratio of the channels (nuzzle) and height of the main channel at both sides of the junction. Results revealed that growing Bo number altered flow regime while viscosity ratio did not have any considerable effect. Incline in width of the nuzzle generated small droplet whereas its increment enlarged droplet size. Inequality in the height of the main channel at the junction postponed droplet formation but generated larger one.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.