Abstract

Unique and remarkable interferences were observed when dried blood spot (DBS) sampling was used in conjunction with liquid chromatography/mass spectrometry (LC/MS) assays. In particular, chromatographic retention time shifting and chromatographic peak shape distortion were observed, along with a severe suppression of MS signal intensity. The type of DBS cards, and chromatographic conditions were investigated using the same set of test compounds to gain insight into these interferences. It was determined that a constituent of the DBS cards, primarily sodium dodecyl sulfate (SDS), was responsible for the interferences by means of an ion-pairing mechanism. SDS formed ion pairs with compounds containing basic amine groups, which resulted in increased retention on a C 18 stationary phase, peak shape distortion and ion suppression. These interferences were greatly alleviated and/or completely overcome with non-acidic mobile phases and/or DBS cards with no SDS coating. To the best of the authors’ knowledge, this is the first in-depth report of interferences induced by DBS cards.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.