Abstract

Microwave-assisted sample digestion using closed vessels is becoming the standard for trace analysis because contamination and losses can be better controlled. Gases are generated during digestion and there is an increment of the internal vessel pressure. Consequently, vessels venting may occur depending on the design of the vessel and the maximum pressure it can stand for. In the present work it was observed that it is possible to allow venting during the digestion without losing volatile analytes, such as As, when properly controlling heating and chemical conditions. Recoveries for As in certified reference materials of animal tissues ranged from 94 to 112% despite mass losses as high as 62%mm−1 observed in the digests. However, for Hg(II) in medium containing chlorides recoveries were poor. The efficiency of digestion was measured by determination of organic carbon contents in digests and they ranged from 0.10 to 0.19% for plant and animal tissues. The temperature gradient along the vessel height is important for avoiding losses of volatile elements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.