Abstract

Al was determined in the stabilized temperature platform furnace with very few interferences. No interferences were found for several metal nitrates, sulfates or phosphates, or for NaCl. The Al absorbance signal was delayed in the presence of MgCl 2 but there was no interference. This led to the use of 50 μg Mg(NO 3) 2 as a matrix modifier for Al. There were no interferences for CaCl 2 but it was particularly important to use new pyrolytically coated tubes to avoid “aging” effects. CuCl 2 provided a very persistent interference that was reduced when the Mg(NO 3) 2 matrix modifier was used that permitted a char temperature of 1700°C. Perchloric acid interferences were severe with improperly coated graphite tubes but did not exist up to 0.5 M HClO 4 when the new pyrolytically coated tubes were used. A serum Al method was tested briefly and no problems were found. Al was determined in seawater with no influence from the salinity of the sample and less than 0.6 μg/1 Al in seawater could be detected.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.