Abstract

The preliminary design studies for developing a prompt-gamma rays in vivo neutron activation analysis facility are in construction at Neutron Activation Research Centre of Ferdowsi University of Mashhad. In this work, activation rate distribution is studied in different bilateral configurations in order to reduce the CV of activation rate distribution throughout the body to less than 10%. It means that the maximum RMS deviation from the mean value of distribution must be smaller than 0.5%. The best case includes two pairs of (241)Am-Be neutron sources. Each pair (with 70 cm distance between two sources) is positioned within up and down 40 x 110 cm(2) area graphite collimators. Also, four moderator/reflector objects are added to transversal body surfaces with 5 cm thick water and 3 cm thick graphite. The proper thickness of polyethylene pre-moderators is chosen 8 mm. The RMS deviation of distribution is 0.3% with the mean activation rate of 1.85 x 10(-6) cm(-3).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.