Abstract

A two-coordinate detector based on a 10B layer and a gas-discharge chamber with a sensitive area of 128 mm2 and 3-mm-thick input and output windows intended for recording slow neutrons are investigated. The recording efficiency is more than 3% at a wavelength of 1.82 A. The recording efficiency is estimated to be ~8% at a wavelength of 8 A. The measured background contribution is less than 0.0001% of the thermal neutron one. For the Ar + 25%CO2 + 0.3%CF3Br gas mixture, the spatial resolution is about 2 mm along one of the coordinates under standard conditions. The neutron position is determined by the charge division method.