Abstract
Korea Research Institute of Standards and Science (KRISS) has assembled a new flow-control system (FCS-705) for the study of advanced applications and calibration of different types of vacuum gauges by comparison method. This report details the measurement of the outgassing rate of the main vacuum chamber by the pressure rate of rise method. Constant pressure was generated inside the vacuum chamber by employing a dynamic flow control method. Nitrogen gas flow data was recorded to demonstrate that constant pressure was controlled in the vacuum chamber. The measurements were performed in a sequence of increasing pressures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.