Abstract

Objective To study the spectrum structure of X-ray and simplify the description method of it. Methods By the general program of Monte Carlo Geant4 to study X-ray spectrum structure and angular distribution of X-ray energy spectrum. Results The results of percentage depth doses and profile curves at different depths of any radiation fields in water phantom calculated by Monte Carlo Geant4 were consistent with the measurements. Conclusions In radiotherapy, X-ray with same nominal energy have very similar dosimetry characteristics, this indicates they have very similar energy spectrum and angular distribution of spectrum. the differences of dosimetric details reflect the different details of the X-ray spectrum structure and angular distribution of X-ray energy spectrum. These studies have very important significances to rapidly build precise virtual source modeling for Monte Carlo calculation based on dose curves measurements in water phantom. Key words: Linear accelerators; X-ray energy spectrum; Angular distribution; Monte Carlo

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.