Abstract

Pedicle screw loosening at the bone–screw interface is the main complication in pedicle screw fixation. The transpedicular polymethylmethacrylate screw augmentation technique has recently become the general technique used to re-operatively overcome pedicle screw loosening. This study investigates the fatigue resistance of PMMA bone cement augmentation. Twenty-seven porcine thoracic vertebral bodies were collected. The BMD was measured using dual X-ray absorptiometry. Each vertebral body was instrumented with one pedicle screw and mounted in a material testing system. Fatigue testing was performed by implementing a cranio-caudal sinusoidal, cyclic (5[Formula: see text]Hz) load. This study shows that transpedicular pedicle screw augmentation with PMMA exhibits similar fatigue resistance as traditional pedicle screw implantation. However, in histomophometrical analysis, the transpedicular pedicle screw augmentation with PMMA has greater anti-deformation capacity than traditional pedicle screw implantation. Transpedicular pedicle screw augmentation with PMMA improves more screw holding power to prevent pedicle screw loosening.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.