Abstract

There are clinical reports of explosions occurring in glyceryl trinitrate (GTN) patches left on patients who are defibrillated and it is widely recommended that such patches are removed before attempting defibrillation. We devised an experimental set up that simulated the electrical conditions under which human defibrillation is conducted, thus making possible accurate measurements of defibrillator wave form. No problems occurred with GTN patches that incorporated a plastic backing. However, with GTN patches that incorporated a metal mesh backing, an explosive effect was consistently observed when current from the defibrillator electrode flowed through the metallic backing. The GTN paste in the patches was unaffected and therefore, we conclude that the apparent explosions reported are solely due to voltage breakdown as the paddle voltage is close to 3 kV resulting from the 360 J of energy. Our results emphasise the need for good technique during defibrillation to avoid current flow through such patches. In addition, our results have implications for the design of patches used for the transcutaneous delivery of drugs.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.