Abstract

It has been previously reported that thermal emissive bands (TEB) 27–29 in the Terra (T-) MODerate resolution Imaging Spectroradiometer (MODIS) have been significantly affected by electronic crosstalk. Successful linear theory of the electronic crosstalk effect was formulated, and it successfully characterized the effect via the use of lunar observations as viable inputs. In this paper, we report the successful characterization and mitigation of the electronic crosstalk for T-MODIS band 30 using the same characterization methodology. Though the phenomena of the electronic crosstalk have been well documented in previous works, the novel for band 30 is the need to also apply electronic crosstalk correction to the non-linear term in the calibration coefficient. The lack of this necessity in early works thus demonstrates the distinct difference of band 30, and, yet, in the same instances, the overall correctness of the characterization formulation. For proper result, the crosstalk correction is applied to the band 30 calibration coefficients including the non-linear term, and also to the earth view radiance. We demonstrate that the crosstalk correction achieves a long-term radiometric correction of approximately 1.5 K for desert targets and 1.0 K for ocean scenes. Significant striping removal in the Baja Peninsula earth view imagery is also demonstrated due to the successful amelioration of detector differences caused by the crosstalk effect. Similarly significant improvement in detector difference is shown for the selected ocean and desert targets over the entire mission history. In particular, band 30 detector 8, which has been flagged as “out of family” is restored by the removal of the crosstalk contamination. With the correction achieved, the science applications based on band 30 can be significantly improved. The linear formulation, the characterization methodology, and the crosstalk effect correction coefficients derived using lunar observations are once again demonstrated to work remarkably well.

Highlights

  • Terra (T-) MODerate resolution Imaging Spectroradiometer (MODIS) is a legacy Earth remote sensing instrument that has completed over 15 years of effective on orbit flight, capturing various geophysical changes of Earth in a broad range of spectral wavelengths

  • T-MODIS band 30 is one of the four photovoltaic (PV) long wave infrared (LWIR) bands that are contaminated by electronic crosstalk [3]

  • In our previous works [3,4,5,6,7,8,9,10,11], we have investigated and reported the electronic crosstalk phenomena in the T-MODIS PV LWIR bands 27–29 and the mitigation of the the electronic crosstalk phenomena in the T-MODIS PV LWIR bands 27–29 and the mitigation of the same in the Level-1B (L1B) product

Read more

Summary

Introduction

Terra (T-) MODerate resolution Imaging Spectroradiometer (MODIS) is a legacy Earth remote sensing instrument that has completed over 15 years of effective on orbit flight, capturing various geophysical changes of Earth in a broad range of spectral wavelengths. The MODIS instrument is a cross track scanning radiometer orbiting the Earth on a sun synchronous polar orbit with an altitude of approximately 705 km [1]. The spectral range varies from the visible blue wavelengths from about. Of spatial over 36 over segregated spectral spectral channelschannels known spectral. Terms of resolution, data areincaptured in m, 250and m, 500 and 1 respectively. In order to achievetothe traceability to groundMODIS based references, resolution.

30. T-MODIS
Correction Algorithms and Coefficients
Crosstalk
It is quite evident from Figure
Crosstalk Coefficients
Crosstalk Correction in BB Calibration
Impact on Non-Linear Coefficients
Impact on Linear Term
Crosstalk Correction in L1B EV Products
EV Imagery
K toto3the
30. Figure
10. Crosstalk
16. Terra band3030 detector difference at Libya
Conclusions