Abstract

AbstractPurposeThis study investigates three different calibration methods for the selection of background pixel intensity.MethodsFilm‐by‐Film (FBF) Method: Each film serves as its own control. Batch‐by‐Film (BBF) Method: A single film is used as a control for all calibration films. Generic (GEN) Method: A generic value (65535) is used as the background pixel value for all calibration films.Three calibration curves were established for the red, green, blue, and RGB channels, and the Radbard NIH (image) curve‐fitting model was used to predict the dose. Sensitivity at different dose levels was quantified by calculating the first derivative of each color channel.ResultsThe GEN method exhibited a difference of up to 6% between the predicted and delivered doses below 2 Gy. The changes in optical density when using the GEN method differed significantly (p<0.0001) from those of the FBF and BBF methods. In the dose range 5–30 Gy, the percentage difference between the predicted and delivered doses for the FBF, BBF, and GEN methods was within 2%. Both the red and green channels demonstrated higher sensitivity than the blue channel over the dose range of 2–30 Gy.ConclusionsThe FBF method is more accurate than the BBF and GEN methods because it accounts for inter‐film variations. The Radbard NIH (image) curve‐fitting function proved suitable for predicting the dose for all the three calibration methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.