Abstract

Aspiration of subglottic secretions is a widely used intervention for prevention of ventilator-associated pneumonia. However, using the Hi-Lo Evac endotracheal tube (Hi-Lo Evac; Mallinckrodt; Athlone, Ireland) (Evac ETT), dysfunction of the suction lumen and subsequent failure to aspirate the subglottic secretions are common. Our objective in this study was to determine the causes of suction lumen dysfunction experienced with the Evac ETT. We studied 40 adult patients intubated with the Evac ETT. In all cases for which dysfunction of the suction lumen was observed, the subglottic suction port was examined visually using a flexible bronchoscope. Dysfunction of the suction lumen occurred in 19 of 40 patients (48%). In 17 of these (43%), it was attributed to blockage of the subglottic suction port by suctioned tracheal mucosa. Evacuation of subglottic secretions using the Evac ETT is often ineffective due to prolapse of tracheal mucosa into the subglottic suction port.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.