Abstract

BackgroundThe design of an external fixator with the optimal biomechanical function and the lowest profile has been highly pursued, as fracture healing is dependent on the stability and durability of fixation, and a low profile is more desired by patients. The plate-type external fixator, a novel prototype of an external tibial fixation device, is a low profile construct. However, its biomechanical properties remain unclear. The objective of this study was to investigate the stiffness and strength of the plate-type external fixator and the unilateral external fixator. We hypothesized that the plate-type external fixator could provide higher stiffness while retaining sufficient strength.MethodsFifty-four cadaver tibias underwent a standardized midshaft osteotomy to create a fracture gap model to simulate a comminuted diaphyseal fracture. All specimens were randomly divided into three groups of eighteen specimens each and stabilized with either a unilateral external fixator or two configurations of the plate-type external fixator. Six specimens of each configuration were tested to determine fixation stiffness in axial compression, four-point bending, and torsion, respectively. Afterwards, dynamic loading until failure was performed in each loading mode to determine the construct strength and failure mode.ResultsThe plate-type external fixator provided higher stiffness and strength than the traditional unilateral external fixator. The highest biomechanics were observed for the classical plate-type external fixator, closely followed by the extended plate-type external fixator.ConclusionsThe plate-type external fixator is stiffer and stronger than the traditional unilateral external fixator under axial compression, four-point bending and torsion loading conditions.

Highlights

  • The design of an external fixator with the optimal biomechanical function and the lowest profile has been highly pursued, as fracture healing is dependent on the stability and durability of fixation, and a low profile is more desired by patients

  • Traditionally, external fixators have been selected as osteosynthesis devices for the treatment of open tibial fractures and certain closed tibial fractures with severe injury to soft tissue [1, 2]

  • Previous studies have demonstrated that the stiffness of a fixation device is a principal determinant of interfragmentary movement, which has a significant effect on the mechanism and progression of fracture healing [5, 6]

Read more

Summary

Introduction

The design of an external fixator with the optimal biomechanical function and the lowest profile has been highly pursued, as fracture healing is dependent on the stability and durability of fixation, and a low profile is more desired by patients. We hypothesized that the plate-type external fixator could provide higher stiffness while retaining sufficient strength. External fixation devices provide a promising and satisfactory alternative for better soft tissue care and for preserving periosteal perfusion to the regions of fracture [3, 4]. They can be selected for use as interim or definite devices of fracture fixation [4]. The large profile of the implants, which tends to confer inconvenience to patients during dressing and

Objectives
Methods
Results
Discussion
Conclusion