Abstract

Venous oxygen saturation reflects venous oxygenation status and can be used to assess treatment and prognosis in critically ill patients. A novel method that can measure central venous oxygen saturation (ScvO2) non-invasively may be beneficial and has the potential to change the management routine of critically ill patients. The study aims to evaluate the potential of sublingual venous oxygen saturation (SsvO2) to be used in the estimation of ScvO2. We have developed two different approaches to calculate SsvO2. In the first one, near-infrared spectroscopy (NIRS) measurements were performed directly on the sublingual veins. In the second approach, NIRS spectra were acquired from the sublingual tissue apart from the sublingual veins, and arterial oxygen saturation was measured using a pulse oximeter on the fingertip. Twenty-six healthy subjects were included in the study. In the first and second approaches, average SsvO2 values were 75.0% ± 1.8 and 75.8% ± 2.1, respectively. The results of the two different approaches were close to each other and similar to ScvO2 of healthy persons (> 70%). Oxygen saturation of sublingual veins has the potential to be used in intensive care units, non-invasively and in real-time, to estimate ScvO2.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call