Abstract

Background and objectiveExcessive noise in the neonatal intensive care unit (NICU) may lead to serious long-term effects on hearing and sensory development in newborns. As such, the maximum allowed noise level is 45 A-weighted decibels (dBA). Studies regarding noise exposure to ventilated preterm infants show inconsistent results; however, these studies also vary considerably in their methodology in terms of noise ascertainment. We hypothesized that the study methodology can significantly influence data quality when measuring noise levels. In this study, we aimed to investigate whether the variations in ventilator noise levels in NICUs could be a result of methodological differences in study designs.MethodsA ventilator circuit was set up using nasal continuous positive airway pressure (nCPAP) and high-frequency (HF) modes with nasal prongs. Noise levels were measured using a commercially calibrated noise meter. Three different scenarios were tested: (1) measurements were taken at different angles (0° to 180°), with 180° facing the end of the nasal prongs, without a mannequin, with the membrane/orifice of the noise meter placed 2 mm laterally from the prongs; (2) noise levels were measured at 180° at distances of 0-20 mm from the nasal prongs; (3) measurements were taken in the oral cavity of a life-size intubation mannequin of a newborn baby.ResultsOverall, the noise levels produced at different settings varied significantly, ranging from 45.7 dB to 82.2 dB. The average environmental background noise was 44.4 dB. Noise levels typically increased as the angle increased, with the highest noise level recorded at 180° for both HF and nCPAP modes, at 58.4 dB and 58.2 dB, respectively. Noise levels recorded at HF were slightly higher than nCPAP values. Furthermore, with regard to distance, the highest mean value, 82.2 dB, was recorded with the noise meter approximately 3 mm from the nasal prongs, and the lowest mean value, 47.6 dB, was recorded at ~20 mm. During trials with the mannequin, the lowest value, 50.1 dB, was recorded at the entrance of the mouth with slightly higher values being recorded within the oral cavity.ConclusionThe results indicate that small changes in experimental settings, such as positioning and distance from the nasal prongs, can greatly influence noise levels, particularly above the recommended levels for neonates. These differences may be attributed to wind-generated noise. In summary, some study results are potentially influenced more by the study design than the device type or ventilator setting. We recommend further research and detailed reporting in the NICU to gain deeper insights into the topic.

Highlights

  • Infants in neonatal intensive care units (NICUs) are regularly exposed to noise from people conversing, and devices, such as alarms and ventilators

  • Three different scenarios were tested: (1) measurements were taken at different angles (0° to 180°), with 180° facing the end of the nasal prongs, without a mannequin, with the membrane/orifice of the noise meter placed 2 mm laterally from the prongs; (2) noise levels were measured at 180° at distances of 0-20 mm from the nasal prongs; (3) measurements were taken in the oral cavity of a life-size intubation mannequin of a newborn baby

  • Noise levels typically increased as the angle increased, with the highest noise level recorded at 180° for both HF and nasal continuous positive airway pressure (nCPAP) modes, at 58.4 dB and 58.2 dB, respectively

Read more

Summary

Introduction

Infants in neonatal intensive care units (NICUs) are regularly exposed to noise from people conversing, and devices, such as alarms and ventilators. While many studies have demonstrated that ventilator noise exceeds recommended levels, there are still discrepancies in reported results, which may be due to differences in experimental settings. Excessive noise in the neonatal intensive care unit (NICU) may lead to serious long-term effects on hearing and sensory development in newborns. Studies regarding noise exposure to ventilated preterm infants show inconsistent results; these studies vary considerably in their methodology in terms of noise ascertainment. We aimed to investigate whether the variations in ventilator noise levels in NICUs could be a result of methodological differences in study designs

Objectives
Methods
Results
Discussion
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.