Abstract

This paper shows how experiments with electrical discharges from the 18th century onward led to their use as sources in atomic spectroscopy and how the invention of the annular inductively coupled plasma (ICP) some 30 years ago arose from the need to solve a problem that necessitated the use of a high-temperature source. The search for such a source followed a fairly logical pattern involving dc plasma jets and an ICP such as had been used by T. B. Reed for crystal growing. The ellipsoidal plasma used by Reed was not entirely suitable as a spectroscopic source, since the analytical sample either mixed with the plasma gases or passed around the plasma, resulting in matrix effects and a diminution in the emission. It is shown how suitable modification of the plasma torch with attention to gas flows made it possible to produce an annular or tunnel plasma through which the sample aerosol could be passed, resulting in an annular ICP with greatly improved spectroscopic properties. The further refinements to the source and ancillary equipment are also discussed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.