Abstract
Introduction. An experimental epidural hematoma model was used to study the relation of ultrasound indices, namely, transcranial color-coded-Doppler (TCCD) derived pulsatility index (PI), optic nerve sheath diameter (ONSD), and pupil constriction velocity (V) which was derived from a consensual sonographic pupillary light reflex (PLR) test with invasive intracranial pressure (ICP) measurements. Material and Methods. Twenty rabbits participated in the study. An intraparenchymal ICP catheter and a 5F Swan-Ganz catheter (SG) for the hematoma reproduction were used. We successively introduced 0.1 mL increments of autologous blood into the SG until the Cushing reaction occurred. Synchronous ICP and ultrasound measurements were performed accordingly. Results. A constant increase of PI and ONSD and a decrease of V values were observed with increased ICP values. The relationship between the ultrasound variables and ICP was exponential; thus curved prediction equations of ICP were used. PI, ONSD, and V were significantly correlated with ICP (r 2 = 0.84 ± 0.076, r 2 = 0.62 ± 0.119, and r 2 = 0.78 ± 0.09, resp. (all P < 0.001)). Conclusion. Although statistically significant prediction models of ICP were derived from ultrasound indices, the exponential relationship between the parameters underpins that results should be interpreted with caution and in the current experimental context.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.