Abstract

At Taiwan Photon Source (TPS), the main liquid-nitrogen (LN2) transfer line of length 600 m for beamline endstations was installed in 2015. It formerly supplied LN2 to maximum 24 beamline endstations. Beamline endstation 13A, of which the aim is advanced and general studies of biological structures and structural kinetics in solution or condensed forms under environmental simulations, from atomic to micrometre scale and time resolution from microsecond to minute, was installed during 2018 and 2019. We designed and self-manufactured one LN2 transfer line according to the requirement of beamline endstation 13A, to supply LN2 into the double-crystal monochromator (DCM), to solve the problem of thermal deformation of the crystal. Otherwise the keep-full device was designed to replace the phase separator of the chiller. In this paper we present the pipeline design, performance testing, keep-full device, pressure reducer, noise reduction and heat-load measurement.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call