Abstract
We require voltage references with a relative output voltage noise of at most 10-6/radicHz in the frequency range of 0.1 mHz-1 Hz. We have investigated the output voltage noise and the external noise sources of a selection of references. Two external noise sources, i.e., supply voltage and temperature, and their coupling to the output voltage of the references were studied. The effect of supply voltage noise was negligible, whereas temperature noise had to be suppressed by stabilization. We found that samples of AD587LN- and MAX6126AASA50-type references fulfilled the requirements if the ambient temperature was stabilized to a level of approximately 50 mK/radicHz.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.