Abstract

The aim of this study is to direct attention to the specific load transfer characteristics of keel-design total disc arthroplasty (TDA) implants that may be underreported. A variety of implants for lumbar TDA are available on the market. One of the main differences between the design types of lumbar TDA implants is whether they use a keel or small spikes/ridges in order to achieve primary stability. The consequences of such design features on load transfer have not been adequately discussed. We report and discuss a case in which new intravertebral bone trabecula have appeared after double-level implantation of a keel-based TDA. We think that the mid- and long-term follow-up radiographs of patients after TDA with keel-design implants should be examined for the presence or absence of such changes. Should our case turn out to be not a singular occurrence, this might have an impact on the design of future TDA implants.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.