Abstract
To help elucidate the mechanism of excimer laser coronary angioplasty (ELCA), intravascular ultrasound (IVUS) imaging was performed in 19 of 29 patients who were treated with ELCA. The results were compared with a non-randomized control group of 18 patients who had IVUS studies both before and after PTCA alone. After ELCA alone, lumen diameter (1.9 x 1.7 mm) and lumen cross-sectional area (CSA) (2.9 mm2) by IVUS were not significantly different from baseline values in the patients before PTCA alone (2.1 x 1.8 mm, 3.2 mm2). After balloon dilatation in the laser treated group, lumen diameter (2.5 x 2.1 mm) and lumen CSA (4.9 mm2) were significantly greater than those post ELCA alone. However, there was no difference in lumen CSA or atheroma CSA in the group treated with excimer laser plus balloon dilatation vs. these measurements in the group treated with PTCA alone. ELCA does not ablate a large amount of atheroma (9% reduction) but creates a pathway to permit easier passage of a PTCA balloon. These quantitative and morphologic results may help explain why the restenosis rate with ELCA is similar to PTCA alone.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.