Abstract

We demonstrate intravascular optical coherence tomography (OCT) imaging with frame rate up to 3.2 kHz (192,000 rpm scanning). This was achieved by using a custom-built catheter in which the circumferential scanning was actuated by a 1.0 mm diameter synchronous motor. The OCT system, with an imaging depth of 3.7 mm (in air), is based on a Fourier domain mode locked laser operating at an A-line rate of 1.6 MHz. The diameter of the catheter is 1.1 mm at the tip. Ex vivo images of human coronary artery (78.4 mm length) were acquired at a pullback speed of 100 mm/s. True 3D volumetric imaging of the entire artery, with dense and isotropic sampling in all dimensions, was performed in <1 second acquisition time.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.