Abstract

To investigate the effects of gaps between the wire pitches of a shape memory Nitinol alloy intravascular endoprosthesis (IVEP), 42 stents with and without gaps were implanted transluminally into the infrarenal abdominal aortae of 11 normal and 10 atherosclerotic rabbits. Digital subtraction angiography was done every 4 weeks to examine the restenosis rate and major side branch patency. Restenosis rates within IVEPs with gaps were significantly lower than in stents without gaps in both normal and atherosclerotic rabbits. The side branch patency rate was significantly higher with IVEPs having gaps. These results suggest that the placement of gaps between wire pitches reduced the restenosis rate within IVEPs, and prevented the obstruction of major side branches.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.