Abstract

To evaluate the intrarater reliability for examining active range of motion (ROM) and isometric strength of the shoulder and elbow among asymptomatic female team handball athletes and a control group using a manual goniometer and hand-held dynamometry (HHD). 22 female team handball athletes (age: 21.0 ± 3.7 years) and 25 volunteers (13 female, 12 male, age: 21.9 ± 1.24 years) participated to determine bilateral ROM for shoulder rotation and elbow flexion/extension, as well as isometric shoulder rotation and elbow flexion/extension strength. Subjects were assessed on two separate test sessions with 7 days between sessions. Relative (intraclass correlation coefficients (ICC) and standard error of measurement (SEM) reliability were calculated. Reliability for ROM and strength were good to excellent for both shoulders and groups (athletes: ICC = 0.94-0.97, SEM 1.07°-4.76 N, controls: ICC = 0.96-1.00, SEM = 0.00 N-4.48 N). Elbow measurements for both groups also showed good-to-excellent reliability (athletes: ICC = 0.79-0.97, SEM = 0.98°-5.94 N, controls: ICC = 0.87-1.00, SEM = 0.00 N-5.43 N). It is important to be able to reliably reproduce active ROM and isometric strength evaluations. Using a standardized testing position, goniometry and HHD are reliable instruments in the assessment of shoulder and elbow joint performance testing. We showed good-to-excellent reproducible results for male and female control subjects and female handball athletes, although the single parameters in ROM and strength were different for each group and between the shoulders and elbows.