Abstract

BackgroundThis was a prospective surveillance study to investigate reports on the safety and frequency of use of intraosseous (IO) access in neonates.MethodsOver a two-year period, paediatric hospitals in Germany were asked to report all cases of IO access to the nationwide Surveillance Unit for Rare Paediatric Diseases (ESPED). Hospitals reporting a case submitted responses via an anonymised electronic questionnaire, providing details on indication, success rate, system used, location, duration to first successful IO access, complications, alternative access attempts and short-term outcome. We present a subset of data for IO use in infants of less than 28 days.ResultsA total of 161 neonates (145 term and 16 preterm born infants) with 206 IO access attempts were reported. In 146 neonates (91%), IO access was successfully established, and success was achieved with the first attempt in 109 neonates (75%). There was no significant impact of gestational age or provider’s educational level on success rates. In 71 infants with successful IO access (79%), the estimated duration of placement was less than 3 min. The proximal tibia was the predominant site used. A semiautomatic battery-driven device was used in 162 attempts (88%). The most often applied medications via IO access were crystalloid fluid and adrenaline. Potentially severe complications occurred in 9 patients (6%).ConclusionWithin this surveillance study, IO access in neonates was feasible and safe. IO access is an important alternative for vascular access in neonates.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call