Abstract
The mobile SIREMOBIL Iso-C(3D) C-arm (Siemens, Erlangen, Germany) is the first device permitting intraoperative, three-dimensional representation of bone structures. A high-resolution, isotropic 3D data cube in the isocenter with sides of approximately 12 cm is calculated simultaneously. The SIREMOBIL Iso-C(3D) is linked to the navigation system. This makes it possible to transfer the generated 3D data directly to the linked navigation system without the need for surgeon-dependent registration. In this prospective clinical trial, we evaluated the accuracy of pedicle screw placement using this device. In 61 patients, a total of 302 pedicle screws were placed. Only in five cases (1.7%) were misplacements of > or =2 mm shown in postoperative control CT. The average fluoroscopy time was 1.28+/-0.56 min, and the average operative duration was 103.26+/-23.3 min. There were no postoperative neurological complications in any of the 30 patients. From these data, we conclude that Iso-C(3D) navigation is a very accurate method for the placement of pedicle screws.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Der Chirurg; Zeitschrift fur alle Gebiete der operativen Medizen
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.