Abstract

This study assesses the feasibility of real-time surgical navigation to plan and guide sequential steps during mandible reconstruction on a series of cadaveric specimens. An image-guided surgical (IGS) system was designed including customized mandible and fibula fixation devices with navigation reference frames and an accompanied image-guided software. The mandibular and fibular segmental osteotomies were performed using the IGS in all five cadaveric patients. Procedural time and cephalometric measurements were recorded. Five real-time IGS mandibulectomy and fibular reconstruction were successfully performed. The mean Dice score and Hausdorff-95 distance between the planned and actual mandible reconstructions was 0.8 ± 0.08 and 7.29 ± 4.81 mm, respectively. Intercoronoid width, interangle width, and mandible projection differences were 1.15 ± 1.17 mm, 0.9 ± 0.56 mm, and 1.47 ± 1.62 mm, respectively. This study presents the first demonstration of a comprehensive image-guided workflow for mandibulectomy and fibular flap reconstruction on cadaveric specimens and resulted in adequate cephalometric accuracy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.