Abstract

A prospective study was lead in order to analyze the accuracy of an X-ray device settled in the operating room for margin assessment, when performing breast-conserving surgery. One hundred and seventy patients were included. All lesions were visible on the preoperative mammograms. An intraoperative X-ray of the lumpectomy specimen was systematically performed for margins assessment. Final histological data were collected and the accuracy of intraoperative specimen radiography (IOSR) for margin assessment was analyzed. IOSR allowed an evaluation of margins status in 155 cases (91.2%). After final histological examination, the positive margins rate would have been 6.5% if margin assessment had relied only on IOSR. Margin assessment with a two-dimensional X-ray device would have allowed the achievement of negative margins in 93.5% of the cases. Moreover, this procedure allows important time-saving and could have a substantial economical impact.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.