Abstract
A clinical method to localize the recurrent laryngeal nerve intraoperatively in order to minimize the risk of accidental injury is presented. By means of an electrode, the nerve was stimulated by a pulsed contact current. The resulting muscle potential was detected using an electrode placed in the larynx. We applied this technique during the time period between 1 January 1997 and 31 December 1998. In 96 cases of primary thyroid resection, the recurrent nerve was identified 167 times. The patients were operated on for nodular thyroid (n=85), Grave's disease (n=9) and malignant papillary goiter (n=2). Retrospectively, the rate of intraoperative nerve injury, equivalent to the rate of postoperative transient and permanent nerve palsy, was 1.04% in the 96 patients and 0.60% with respect to the 167 nerves at risk. The rate of failure of the method was 7.29%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.