Abstract

AimsThis trial was designed to test the hypothesis that shock efficacy during follow-up is not impaired in patients implanted without defibrillation (DF) testing during first implantable cardioverter-defibrillator (ICD) implantation.Methods and resultsBetween February 2011 and July 2013, 1077 patients were randomly assigned (1 : 1) to first time ICD implantation with (n = 540) or without (n = 537) DF testing. The intra-operative DF testing was standardized across all participating centres, and all ICD shocks were programmed to 40 J irrespective of DF test results. The primary end point was the average first shock efficacy (FSE) for all true ventricular tachycardia and fibrillation (VT/VF) episodes during follow-up. The secondary end points included procedural data, serious adverse events, and mortality. During a median follow-up of 22.8 months, the model-based FSE was found to be non-inferior in patients with an ICD implanted without a DF test, with a difference in FSE of 3.0% in favour of the no DF test [confidence interval (CI) −3.0 to 9.0%, Pnon-inferiority <0.001 for the pre-defined non-inferiority margin of −10%). A total of 112 procedure-related serious adverse events occurred within 30 days in 94 patients (17.6%) tested compared with 89 events in 74 patients (13.9%) not tested (P = 0.095).ConclusionDefibrillation efficacy during follow-up is not inferior in patients with a 40 J ICD implanted without DF testing. Defibrillation testing during first time ICD implantation should no longer be recommended for routine left-sided ICD implantation.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call