Abstract

The intraoperative quality assessment of the arteriovenous fistula for hemodialysis is an essential process to limit early failure due to technical problems or inadequate vascular quality. This step is not clearly defined in the literature with no recommendations. We selected published articles related to the topic of intraoperative quality control of the vascular access for hemodialysis. The intraoperative blood flow measurement greater than 120 ml/min in autologous fistula and less than 320 ml/min in arteriovenous graft was described as predictive factors for early failure. The blood flow measurement should be performed after the confection of the anastomosis. When blood flow is limited, fistulography is an essential step to assess patency.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.