Abstract

Transoesophageal Doppler cardiac output measurement was evaluated against the thermodilution method in eleven patients undergoing elective cardiac surgery. A total of 106 pairs of Doppler and thermodilution values were obtained. Cardiac output was measured over a range of 2.3 l.min-1. to 11.51 l.min-1. The mean difference between the Doppler and thermodilution measures was -1.0 l.min-1 (thermodilution minus Doppler). Standard deviction was 1.8 l.min-1. This is a significant difference. It is concluded that these techniques do not agree when measuring cardiac outputs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.