Abstract

Aspiration around uncuffed endotracheal tubes in paediatric patients is an uncommon event. Uncuffed endotracheal tubes are commonly used in paediatrics even when the risk of gastric aspiration is significant. A case history is presented of an 81/2-year-old who aspirated during a laparotomy for bowel obstruction. The patient was intubated with an uncuffed #6.0 Portex endotracheal tube. A small leak was noted during positive pressure ventilation. The aspiration was significant such that the patient required 24 hours of positive end-expiratory pressure (PEEP) in an intensive care unit. Uncuffed endotracheal tubes are used because post-intubation stridor is minimized, the lungs are compliant and the funnel-shaped trachea is more likely to secure the airway from aspiration. An uncuffed tube however, does not completely seal the airway, particularly in the presence of an audible "leak." As a result of this case, we now consider the use of a cuffed endotracheal tube in children whenever a 5.5 internal diameter or larger is required. When the risk of pulmonary aspiration is present, the cuff is inflated to reduce the likelihood of contamination of the airway.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.