Abstract

To examine the feasibility of the use of a flexible intraocular rake to remove epiretinal membranes (ERM). Interventional case series. A multi-tine flexible intraocular rake was designed. Ten consecutive patients with idiopathic epiretinal membranes underwent vitrectomy and removal of the ERM with the intraocular rake. Outcome measures included intraoperative complications, Early Treatment Diabetic Retinopathy Study visual acuity, rate of recurrent ERM, and rate of persistent cystoid macular edema. The rake was used successfully to remove the ERM in all cases. Two patients had intraoperative retinal tears. At a mean postoperative follow-up examination of 24 weeks, the visual acuity was improved significantly (20/90 before the operation vs 20/50 after the operation; P<.05). One patient had a recurrent epiretinal membrane. The flexible intraocular rake appears to be capable of the removal of idiopathic ERM. The increased tolerance for retinal contact that results from the flexible tines may make the rake of value to many vitreoretinal surgeons.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.