Abstract

To evaluate intraocular pressure (IOP) changes during experimental vitrectomy and the efficacy of Constellation Vision System's IOP control (IOPc) feature in reestablishing baseline pressure. Using a pressure transducer in freshly enucleated porcine eyes, a broad range of parameters (baseline pressures, aspiration levels, and cut rates) were tested with 23- and 25-gauge probes and IOPc turned ON versus OFF. IOPc turned ON was significantly more effective than IOPc turned OFF in controlling IOP drop and stabilizing pressure during vitrectomy using a wide range of baseline pressures (20-70 mmHg). The 23-gauge system consistently presented a reduced drop from baseline compared with the 25-gauge system. The overall average drop for the 23- and 25-gauge systems was 12.79 mmHg and 21.17 mmHg, respectively. Both gauge sizes reestablished baseline pressure approximately 1.6 seconds after the initial pressure drop generated at the beginning of aspiration. A peak of IOP (overshooting) was observed when the pressure was returning to baseline using both 23- and 25-gauge systems. Using IOPc feature turned ON, 23- and 25-gauge probes were effective in reestablishing and sustaining baseline infusion pressures, although 23-gauge probes showed less IOP fluctuation than did 25-gauge probes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.