This paper aims to review the current methods available for the measurement of intraocular pressure after myopic laser in situ keratomileusis for the correction of myopia. Searches were performed for studies that assessed or compared various methods of intraocular pressure assessment. There were 20 eligible studies that explored the use of pneumotonometry, pressure phosphene tonometry, rebound tonometry, dynamic contour tonometry, statistical modeling, mathematical formulae, ocular response analyzer and even measuring intraocular pressure on the nasal cornea. Our review shows that an ideal method would be one that is independent of corneal factors. Dynamic contour tonometry and pressure phosphene tonometry held promise in research settings. More studies need to be done to validate the new methods of intraocular pressure assessment, especially in glaucoma patients. It is important to empower laser in situ keratomileusis patients with knowledge of these difficulties and potential implications for the future.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.