Abstract
ObjectivesTo evaluate intraobserver and interobserver agreement of manually encompassed lung lesions for perfusion measurements using volume-perfusion computed tomography (VPCT). Materials and methodsInstitutional review board approval and informed consent were obtained. HIPAA guidelines were followed. A 65-s dynamic study was acquired with scan parameters 80kV, 60mAs (80mAs for patients ≥70kg), 128×0.6mm collimation. Blood flow (BF), blood volume (BV) and Ktrans parameters were determined by syngo volume perfusion CT body with 88 lesions analyzed retrospectively. ResultsWithin-subject coefficients of variation for intraobserver agreement (range 6.59–12.82%) were superior to those for interobserver agreement (range 21.75–38.30%). Size-dependent analysis revealed lower agreements for lesions <4cm as compared to larger lesions. Additionally, agreements of the upper, middle and lower lung zones were different. ConclusionsIntraobserver agreement was substantial for VPCT lung cancer perfusion measurements encouraging the use for tumor characterization and therapy response monitoring. Interobserver agreement is limited and unexperienced readers should be trained before using this new method.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.