Abstract

We evaluated the performance of the Targon PF nailing system in the treatment of complex proximal femoral fractures with a retrospective analysis of prospectively collected data for 94 patients with a complex proximal femoral fracture (subtrochanteric fracture or fracture with reversed fracture line pattern) treated with a Targon PF nail. There were no intra-operative femoral fractures. Late complications requiring reoperation were one femoral fracture and one fracture around the nail secondary to tumour. Seven nails had to be dynamised and a further six 'self-dynamised' at the site of the distal locking screw. One patient required revision of the distal locking. Only one other patient required secondary surgery for a wound haematoma. There were no cases of implant cut-out or non-union. The Targon PF nail represents a progressive development in the design of intramedullary nails for proximal femoral fractures and the results compare favourably with other intramedullary systems used to treat complex proximal femoral fractures.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.