Abstract

Intramedullary instrument/device insertion during total joint replacement can generate high intramedullary pressures, with potential for clinical fat embolism syndrome. We evaluated effects of process parameters on intramedullary pressure generated when hammering an implant into a simulated femur analogue. Lower implant insertion speed, lower hammering force, a rubber- compared with steel-tipped hammer and a larger synthetic bone-to-implant radial gap reduced intramedullary pressure generated. Number of hammer strikes required to insert an implant increased with increasing marrow viscosity, whereas intramedullary pressure decreased with increasing viscosity. Significant intramedullary pressure reduction is obtained by applying more strikes of lower peak force using a rubber hammer tip, as opposed to fewer larger strikes. Caution should be exercised when instrumenting the distal femur, where viscosity is low and very high intramedullary pressures can be generated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.