Abstract
Purpose: To evaluate intrafractional tumor position stability during computed tomography (CT)-guided frameless stereotactic radiation therapy (SRT) for lung or liver cancers, we checked repeated CT scanning, with a fusion of CT and linear accelerator (FOCAL) unit. Methods and Materials: The FOCAL unit is a combination of a linear accelerator (Linac), CT scanner, X-ray simulator (X-S), and carbon table, and is designed to achieve CT-guided SRT with daily CT positioning followed by immediate irradiation while patients keep reduced shallow respirations. To evaluate intrafractional tumor position stability, 50 lung or liver lesions in 20 patients were checked by repeated CT scanning just before and after irradiation, and the obtained images were compared. Results: There was no case with the intrafractional error judged to be greater than 10 mm. In 68% of cases, the intrafractional positioning errors were negligible (0–5 mm). Conclusions: Using the FOCAL unit, SRT for lung or liver cancers could be performed with intrafractional positioning errors not greater than 10 mm.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Radiation Oncology, Biology, Physics
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.