A newly developed infant cranial model shows that accurate, reproducible, and noninvasive measurements of intracranial pressure (ICP) can be made from the anterior fontanelle when fontanelle pressure is referenced from the bony margins adjacent to the fontanelle opening. Also, this model provides insight into the elastic properties of the fontanelle membrane and the pressure/volume relationships governing ICP transmission through the fontanelle window. An anterior fontanelle pressure monitor design based on data from the infant cranial model was used to monitor anterior fontanelle pressure in three infants with elevated ICP and previously inserted ventricular catheters. Measured anterior fontanelle pressure was highly correlated to ICP (r = 0.962) with high reproducibility after blind application and reapplication. Base line adjustment and in situ recalibration were easily achieved, with the monitor showing no sensitivity to patient movement and excellent frequency response.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.