Abstract

The aim of the study was to investigate if laser Doppler perfusion monitoring (LDPM) can be used in order to differentiate between gray and white matter and to what extent microvascular perfusion can be recorded in the deep brain structures during stereotactic neurosurgery. An optical probe constructed to fit in the Leksell® Stereotactic System was used for measurements along the trajectory and in the targets (globus pallidus internus, subthalamic nucleus, zona incerta, thalamus) during the implantation of deep brain stimulation leads (n = 22). The total backscattered light intensity (TLI) reflecting the grayness of the tissue, and the microvascular perfusion were captured at 128 sites. Heartbeat-synchronized pulsations were found at all perfusion recordings. In 6 sites the perfusion was more than 6 times higher than the closest neighbor indicating a possible small vessel structure. TLI was significantly higher (p < 0.005) and the perfusion significantly lower (p < 0.005) in positions identified as white matter in the respective MRI batch. The measurements imply that LDPM has the potential to be used as an intracerebral guidance tool.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.