Abstract
Intracavity optogalvanic detection of 14C utilizes the narrow band specificity of 14CO2 laser resonances in the infrared spectrum, coupled with greatly enhanced sensitivity from amplified effective path lengths in optical cavities. Background (nonresonant) interactions are also enhanced in such cavities, making proper separation of signals from background imperative. Intracavity Optogalvanic Spectroscopy (ICOGS) is similar to cavity ring-down spectroscopy (CRDS) with certain crucial differences. An ICOGS model, based on a basic CO2 laser and optogalvanic effect physics, is presented. Experimental results using a laboratory instrument with a continuous flow of small CO2 samples in a buffer gas illustrate the model. It is shown that acquisition of optimized signal-to-background ratios for the quantitation of 14CO2 in samples of order 10 μg is achievable, considerably smaller than with CRDS.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.