Abstract

Seventy-five patients with various abdominal diseases were studied by intraarterial digital subtraction angiography (IA DSA) during conventional abdominal angiographic examinations. The image quality, tumor stain, and arterial encasements were evaluated and diameters of vessels visualized by IA DSA were measured. The vessels greater than 2.0 mm or more in diameter were equally well demonstrated on IA DSA. IA DSA demonstrated tumor stains and portal system to good advantage. Real time imaging was also advantageous during an embolization procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.