Abstract

BackgroundArthrometers are used to assess knee anterior laxity and to evaluate the integrity of the anterior cruciate ligament. Assessment of knee anterior laxity is crucial part of the clinical examination. The aim of this study was to investigate the intra-rater reliability of the GNRB® in healthy subjects. MethodsIn the study participated 97 subjects and two measurements using a GNRB® arthrometer were performed. Males were tested two times one week a part. Females were tested two times within 24 h. Measurements were performed at the following forces 134 N, 134 N, 150 N, 200 N, and 200 N. To evaluate the reliability of GNRB® measurements an intraclass correlation coefficient was calculated. ResultsIntra-rater reliability of the GNRB® measurements is good for 134 N on the left knee (ICC = 0.848) and the right knee (ICC = 0.788) and for 200 N on the left knee (ICC = 0.805) and the right knee (ICC = 0.756). ConclusionThe GNRB® knee arthrometer has good intra-reliability for measurements at the 134 N and 200 N forces. Reliability can be increased with the standardize position of the subject, the stabilization of the patella and standardize measurement protocol.