Abstract

Background: The application of a cannulated device is a widely used fixation method for hip fractures. Although the breakage of the guide wire during this procedure is an uncommon complication, migration of the broken fragment might be catastrophic. Case Report: We presented a case of intertrochanteric fracture with breakage of the lag screw guide wire in the acetabulum during fixation with cephalomedullary nail. The broken fragment was located with computed tomography (CT) scan and removed through Stoppa approach. Conclusion: We found the Stoppa approach a safe and useful method for extracting the intra-pelvic broken guide wire.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.