Abstract

Permanent implantation of radioactive seeds is a viable and effective therapeutic option widely used today for early-stage prostate cancer. The implant technique has improved considerably during the recent years due to the use of image guidance; however, real-time dose distributions would allow potential cold spots to be assessed and additional seeds added. In this study, we investigate the use of a conventional C-arm fluoroscopy unit for image acquisition and evaluation of dose distribution immediately after the implant. The phantom study indicates that it is possible to obtain seed positions within ±2 mm. A pilot study carried out with three patients indicated that it is possible to obtain seed positions and calculate the dose distribution with C-arm fluoroscopy and about 95% of the seeds were reconstructed within ±2 mm. The results could be further improved with better digital imaging.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.